Home Messtechnik Produkte ACR Systems Inc. IAQ Logger

IAQ Logger

The IAQ Logger, when combined with the Telaire 7001 Handheld CO2 Monitor, records the three leading indicators of indoor air quality: (1) temperature, (2) relative humidity, and (3) carbon IAQ100dioxide. This IAQ solution is ideal for those applictions requiring long-temperature and CO2 on the Telair´s large, easy-to read display.

Prdouct Specifications

  • Size: 107 mm x 74 mm x 22 mm (4.2" x 2.9" x 0.9")
  • Weight: 110 g (3.75 oz)
  • Case Material: Noryl®plastic
  • Mountign: Magnetic backing or looking eyelet
  • Operating Limits: -40 to 70ºC (-40 to 158ºF) and 0 to 95% RH (non-condensing)
  • Clock Accuracy: +/- 2 seconds per day
  • Battery or Power Supply: 3.6 volt Lithium
  • Battery Life: 10 years under normal use
  • No. of Channels: five
  • - One for ambient temperature
  • - One for relative humidity
  • - One for carbon dioxide
  • - Two for temperature and/or RH (with ACR thermistor probes or ACT temp/RH probe)
  • Memory Size: 32 KB (32.67 readings)
  • Sampling Methods:
  • 1. Continuous (First-in, First-out)
  • 2. Stop when full (Fill-then-stop)
  • Sampling Rates: User selectable rates from 8 seconds to once every 5 days
  • Resolution: 8-bit (1 part in 256)
  • PC Requirements: IBM PC or 100% compatible running Windows® 3.1, `95, `98, 2000, Me, Nt or XP with at least 2 MB RAM, 2 MB of hard drive disk space and one free serial port
  • Software Requirements: Trand Reader®Standard

 

TELAIRE 7001

  • Size: 170 mm x 80 mm x 40 mm (6.7" x 3.1" x 1.6")
  • Weight: 280 g (9.9 oz)
  • Case Material: ABS plastic
  • Mounting: Pull-out stand
  • Operating Limits: 0 to 50ºC (32 to 122ºC) and 0 to 95% RH (non-condensing)
  • Power Supply: 6 VDC from external AC/DC adapter (included) or 4 AA Batteries (not included)

 

Sensor Specifications

  • Temperature Sensor (in IAQ Logger):
  • Type: NTC Thermistor
  • Range: -40 to 70ºC (-40 to 158ºF)
  • Accuracy: +/- 0.2ºC over the range of 0 to 70ºC (+/- 0.2ºC over the range of 32 to 158ºF)
  • Resolution: 0.4ºC(0.7ºF) @ 25ºC; better than 1ºC (1.8ºF) between -25 and 70ºC (-13º and 158ºF); better than 2.0ºC (3.6ºF) between -40 and -25ºC (-40ºF and -13ºF)

 

  • RH Sensor (in IAQ Logger):
  • Type: Capacitive thin polymer film
  • Range: 0 to 95% RH
  • Accuracy: +/- 4% RH from 10 to 90º (-20 to 40ºC [-40 ti 104ºF])
  • Resolution: Better than 0.4º RH between 25 and 60º RH at 25ºC (77ºF)

 

  • CO2 Sensor (in Telaire:)
  • Range: 0 to 4,000 ppm voltage output, 0 to 10,000 ppm display
  • Sensitivity: +/- 1 ppm
  • Accuracy: +/- 50 ppm or +/- 5 % reading (whichever is greater)
  • Repeatability: +/- 20 ppm