This line of 12-bit, multi-channel data loggers habe the
same capacitilites as the SmartReader line, but with
greater memory capacity, faster sampling rates, alarm
dial-out and are wireless ready.
Common Specifications
-
○ Size: 107 mm x 74 mm x 22 mm (4.2" x 2.9" x
0.9")
- ○ Weight: 110 g (3.75 oz)
- ○ Case Material: Noryl®plastic
-
○ External Connector: Removable screw-type terminal strip
with Common (-) connection
- ○ Mounting: Magnetic backing or looking eyelet
-
○ Operating Limits: -40 to 70ºC (-40 to 158ºF) an 0 to
95% relative humidity (non-condensing)
- ○ Clock Accuracy: + / - 2 seconds per day
- ○ Battery: 3.6 volt Lithium, 1 Amp-Hour
-
○ Battery-Life: 10 years under normal use (factory
replaceable)
- ○ Power Consumption: 5 to 10 microamps (continuous)
- ○ Memory Size¹: 32 KB (21.5000 readings), 128 KB (87,000 readings], 1.5 MB (1,048,000 readings)
- ¹ See specific models for availability
-
○ Sampling Memory: 1. Continuous (First-in. First-out -
not available with sample rates faster than 8 seconds), 2.
Stop when full (Fill-then-stop)
-
○ Sampling Rates: User selectable rates from 25 per second
(0.04 seconds) to once every 8 hrs¹
-
¹BP-101 Battery Pack or PS-201 Power Supply required for
sample rates faster than 8 seconds (see Accessories section
for dtails)
- ○ Resolution: 12-bit (1 part in 4,,096)
-
○ PC Requirements: IBM PC or 100% compatible running
Windows®3.1. 95, 98, 2000, Me, NT or XP with least 2 MB
RAM, 2 MB of hard drive disk space and one free serial port
- ○ Software Requirements: Trend Reader®Standard
Certifications and Standards:
-
- Certified to CE standard EN50082-1: 1992 ( European
Generic Immunity) covering ESD, PFI, and EFT/B
-
- Meets CE standards EN55011 Class B: 1991 ( European
Radiated Emissions) and CISPR Class B: 1991 ( International
Radiated Emissions)
Resistance to X-Rays:
-
- Tested for protection against a 160kV dose @ 5mA for 30 secondes (150mA-sec) @ 38 inches F.F.D (about 100 times
that of an airport x-ray machine)
-
- Tested for protection against Gamma Ray (equivalent to
0.137-1.38 mega volts) IR 192 - 28 curies @ 30 seond expose
source to object distance 5"
Internal Temperature Sensor (common on all
models, expect SmartReader Plus9)
- ○ Type: Negative Temperature Coefficient Termistor 10,000 ohms @ 25ºC (77ºF)
-
○ Accuracy: + / - 0,2ºC over the range of 0 to 70ºC(+ /
- 0.3ºF over the range of 32 to 158ºF)
- ○ Range: -40º to 70ºC (-40 to 158ºF)
-
○ Resolution: 0.4ºC (0.7ºF) @ 25ºC; better than 1ºC
(1.8ºF) between -25 and 70ºC (-13º and 158ºF); better
than 2.0ºC (3.6ºF) between -40 an -25ºC (-40ºF and
-13ºF)